Essential Information
|
|
Warning— This may not include all manditory upgrades. Please refer to the most current Technical Bulletins for up to date implementation information. |
Panther
Upper Bay
Communication and Organization Processor (COP) Module
The Panther CAN-Bus-Terminator Installation was cut in on production Panther Systems P2242 and above.
| Install Required | Factory Installed | Required for Panther | Required for Fusion |
|---|---|---|---|
| P0000 - P2241 | P2242 and above | Yes | Yes |
Monitor
The Panther LCD Monitor was installed on production Panther Systems P0000 to P1402.
| Factory Installed |
|---|
| P0000 and P1402 |
The Panther LED Monitor was cut in on production Panther Systems P1405 to P2306.
| Factory Installed |
|---|
| P1405 and P2306 |
A New VGA Cable was cut in on production Panther Systems P1589 and above.
| Install on Failure | Factory Installed |
|---|---|
| P0000 - P1589 | P1589 and above |
The Panther PCAP Monitor was cut in on production Panther Systems P2306 and above.
|
Optional Upgrade on Panthers |
Factory Installed | Mandatory Install on Fusions |
|---|---|---|
| P0000 - P2306 | P2307 and above | Yes |
Pipettor Module
Please refer to the most current Technical Bulletins for up to date implementation information.
All Panther PLUS instruments from manufacturing are built with BLDC pipetors.
The BLDC pipettors were cut in on production Panther (only) Systems P03499, 03510, 03511, P03542 and above.
| Factory Installed on Panther ONLY Systems (Classic Panther) |
|---|
| P03499, P03510, P03511, P03542 and above |
|
Note— BLDC Pipettors can ONLY be installed with System Software v7.1 or higher. |
For Panther Systems, the Sample and Reagent Pipettors MUST be of the same type.
For Fusions Systems, the Panther side (Sample and Reagent Pipettors) MUST be of the same type.
The Fusion Pipettor can be either an APP or BLDC Pipettor.
The table below represents the only viable pipettor configurations for standalone Panther or Panther Fusion systems:
| SYSTEM | Sample Pipettor | Reagent Pipettor | Fusion Pipettor |
| Panther | APP | APP | APP or BLDC |
| Panther Fusion | BLDC | BLDC | APP or BLDC |
Pumps Module
A New LCMP Vacuum Seal may be required for system installs at high elevations.
Reagent Bay Module
Please refer to the most current Technical Bulletins for up to date implementation information.
Sample Bay Module
Please refer to the most current Technical Bulletins for up to date implementation information.
Target Capture Reagent (TCR) Module
TCR Mixer Drip Shield was cut in on production Panther Systems P0354 and above.
| Mandatory Install Required on | Factory Installed |
|---|---|
| P0000 - P0353 | P0354 and above |
Molded TCR Carousel was cut in on production Panther Systems P0827 and above.
| Install as needed | Factory Installed |
|---|---|
| P0000 - P0826 | P0827 and above |
TCR
Target capture reagent—An assay-specific reagent added as part of specimen pipetting. Door Laser block was cut in on production Panther Systems P1301 and above.
| Mandatory Install Required on | Factory Installed |
|---|---|
| P0000 - P1300 | P1301 and above |
TCR Door Grounding Brush was cut in on production Panther Systems P1560 and above.
| Mandatory Install Required on | Factory Installed |
|---|---|
| P0000 - P1559 | P1560 and above |
Tip Drawers Module
Please refer to the most current Technical Bulletins for up to date implementation information.
Mid Bay
Chiller Ramp Module
The New Chiller Ramp was cut in on production Panther Systems P0681 and above.
Installation of the New Chiller Ramp requires an exhaust fan and vent. Please click the hyper link for details.
| Install on Failure | Factory Installed |
|---|---|
| P0000 - P0680 | P0681 and above |
Distributor Module
Panther Updated Linear Distributor was cut in on production Panther Systems P02067 - 02076, and P02089 and above.
| Install On Failure | Factory Installed |
|---|---|
| P0000 - P02066 | P02067-P02076, P02098 and above |
Incubators Module
Panther Incubator Weights were cut in on production Panther Systems High Temp-P0655, Transition-P0654, Amp-P0642 and above.
| Install Optional | Factory Installed |
|---|---|
| P0000 - P0641 | P0655 and above |
Panther New Molded Incubator Door with door screw on High Temp was cut in on production Panther Systems P0916 and above.
| For Systems | Factory Installed |
|---|---|
|
P0000 - P0916 Install on Failure If no door screw is present ensure |
P0916 and above |
Panther New Molded Incubator Door with door screw on Trans and Amp were cut in on production Panther Systems P0919 and above.
| Install Optional | Factory Installed |
|---|---|
|
P0000 - P0919 If no door screw is present ensure |
P0919 and above |
Panther Incubator Amp Inc Form Plugs were cut in on production Panther Systems P1251, P2153 and above.
| Mandatory Install if RTFs present | Factory Installed |
|---|---|
| P0000 - P1250 & P1252 | P1253 and above |
Panther New Incubator Shims and Belt were cut in on production Panther Systems P2562 and above.
| Install on Failure | Factory Installed |
|---|---|
| P0000 - P2561 | P2562 and above |
Thermal Sensor Ranges for Aptima Assays
Input Queue Module
Panther New Input Queue was cut in on production Panther Systems P1264 and above.
| Install on Failure | Factory Installed |
|---|---|
| P0000 - P1263 | P1264 and above |
Load Stations Module
Please refer to the most current Technical Bulletins for up to date implementation information.
Luminometer Module
Panther Luminometer Ferrite on PMT
Photomultiplier tube was cut in on production Panther Systems P0169 and P0172 and above.
| Install on Failure | Factory Installed |
|---|---|
| P0000 - P0168 | P0172 and above |
Magnetic Parking Module
Please refer to the most current Technical Bulletins for up to date implementation information.
Magnetic Wash Module
Please refer to the most current Technical Bulletins for up to date implementation information.
Output Queue Module
Please refer to the most current Technical Bulletins for up to date implementation information.
Tip Eject Chute Module
The New Sample Tip Eject Chute was made available for BloodBank systems ONLY.
Lower Bay
Cooling Module
Panther New Cooling Module (Dual Fan) was cut in on production Panther Systems P0734 and above.
| Install as needed | Factory Installed |
|---|---|
| P0000 - P0733 | P0734 and above |
PC
Dell XE2 was cut in on production Panther Systems P1493 and above.
| Install as needed | Factory Installed | Required for Panther | Required for Fusion |
|---|---|---|---|
| P0000 - P1492 | P1493 and above | No | Yes |
Dell XE2 with 16GB RAM
| Factory Installed | Required for Panther | Required for: |
|---|---|---|
| P2671 | No | System SW v7 Panther PLUS Panther Link |
Dell XE2 Minitower was cut into Panther ONLY production on System P3202 and above.
| Factory Installed |
|---|
| P3202 |
Dell XE3 was cut into Panther production as Panther (only) Systems on P03499, 03510, 03511, P03542 and above.
| Factory Installed on Panther ONLY Systems (Classic Panther) |
|---|
| P03499, P03510, P03511, P03542 and above |
The following PCs require a BIOS configuration update before Imaging (Change SATA Operation to RAID ON) .
PC BIOS Settings before Imaging (SATA RAID On)
| PC Part Number | Part Description |
|---|---|
| CMP-01711 / RCMP-01711 | Dell OptiPlex XE2, 16GB |
| CMP-01672 | Dell OptiPlex XE3, 16GB |
Universal Fluid Drawer Module
No Latches can be installed on Systems 001 - 0099 as the frame / chassis will not accommodate the latch hardware.
Hologic Magnetic Latches were cut in on production Panther Systems P1600 and above.
These Hologic latches are on the right (closest to the right side door) as you face the instrument.
| Mandatory Install on | Factory Installed |
|---|---|
| P0100 - P1600 | P1601 - P1937 |
Stratec designed UFD Magnetic Latches were cut in on production Panther Systems P1937 and above.
These Stratec latches are on the left of the UFD as you face the instrument.
| Factory Installed |
|---|
| P1937 and above |
Panther Improved Bottle Caps & Straws were cut in on production Panther Systems P1875 and above.
| Install on Failure | Factory Installed |
|---|---|
| P0000 - P1874 | P1875 and above |
Vacuum System Module
See Panther Vacuum Module page for cut in information.
The Odorcarb Vacuum filter MUST be replaced every 12 months (from the date of install).
Waste Drawer Module
No Latches can be installed on Systems 001 - 0099 as the frame / chassis will not accommodate the latch hardware.
Hologic Magnetic Latches were cut in on production Panther Systems P1600 and above.
These latches are on the left (closest to the drawer interlock) as you face the instrument.
| Mandatory Install on | Factory Installed |
|---|---|
| P0100 - P1600 | P1601 - P1937 |
Stratec designed WD Magnetic Latches were cut in on production Panther Systems P1937 and above.
These Stratec latches are on the right of the WD as you face the instrument.
| Factory Installed |
|---|
| P1937 and above |
Panther Mid-Bay Shields above Waste Drawer were cut in on production Panther Systems P0375 and above.
| Factory Installed |
|---|
| P0375 - 1990 |
Panther Mid-Bay Shields above Waste Drawer with Wider cut out cut in on production Panther Systems P1900 and above.
| Factory Installed |
|---|
| P1900 and above |
Panther Waste Fittings (Polysulfone) were cut in on production Panther Systems P0674 and above.
| Install on Failure | Factory Installed |
|---|---|
| P0000 - P0673 | P0674 and above |
The Odorcarb Vacuum filter MUST be replaced every 12 months (from the date of install).
Fusion
Thermocycler
Vial Cleaning Nests
| Install on | Factory Installed |
|---|---|
| F0000 - F00245 | F0246 and above |
External Component Modules
Chassis
Panther Dual Chassis Fans were cut in on production Panther Systems PXXXX and above.
| Install Optional | Factory Installed |
|---|---|
| P0000 - PXXXX | PXXXX and above |
Panther Track Ready Hole was cut in on production Panther Systems P1937 and above.
| Can be Installed as needed | Factory Installed |
|---|---|
| P0000 - P1936 | P1937 and above |
Panther UDI Tracking Label was cut in on production Panther Systems P2242 and above.
| Install as needed | Factory Installed | Required for Panther | Required for Fusion |
|---|---|---|---|
| P0000 - P2241 | P2242 and above | Yes/No | Yes/No |
Printers
Panther Systems with an HP Officejet Printer may need to have the Auto Shutoff Feature Disabled.
Uninterruptible Power Supply (UPS) Module
ALL Fusion Installs require the Output Voltage on the UPS to be set at 220V.