New Liquid Coupled Membrane (LCM) Pumps for AutoDetect and Oil
What is Affected
- MME-02724 PUMP (OILAD1AD2), LCM 500μL, VAC SEAL is available to be ordered as a spare part.
- MME-02724 can be used to replace LCM pumps that are showing a delayed pullback causing injection failures. These may be reported as PCL1 or PCL2 flags.
- Typically, two pumps should be ordered for any Panther install above 2000ft (610m).
- It is also possible for the oil injection pump to fail the oil process control.