MTU Expansion Installation
To install a PLUS module, the System requires the following:
- Panther Plus can only be installed on Panther Systems Serial #00375 and above.
- PCAP Monitor required if installing MTU
Multi-tube unit—Container used to process tests in the instrument. An MTU contains five separate reaction tubes. The MTU is moved through the instrument by the linear distributor and includes five tiplets for pipettiing to be used in the mag wash station. Expansion ONLY - Dell XE3
- Software v7.x and above
- Waste to Drain - drain must be able to handle a minimum flowrate of 1L/min.
- Waste to Drain - drain must be 15m (49ft) or less from the Waste to Drain pump.
- Waste to Drain - drain must be no higher than 1.2m (3.9ft) from the floor
Parts and Materials Required
- Panther Tool Kit
- Level
- System Software
- Panther Plus v7.1.5 or higher
- Panther Fusion Plus v7.2.1 or higher
- PCAP Monitor Installed
MTU EXPANSION KIT - MTU QUEUE TEACH TOOL [TLS-06904]
- INPUT QUEUE TEACH TOOL [TLS-06359]
- Cart for moving the MTU Expansion Module (Lab Supplied).
Time Required
- 3 Hours
Procedure
The sequence of installation is as follows
- Unboxing MTU Expansion
- Prepare the Panther
- MTU Queue Installation
- Teach the Linear Distributor to MTU Queue
- MTU Sidecar Installation
- Verify alignment in Y-Direction of scissor
- MTU Queue to Sidecar Alignment
- MTU Expansion Autoteach
- Install the Covers