Service Software Verification Procedures
- Verify that the Panther cooling system is working correctly.
- Navigate to the 4. LiVaCo System and Drawer tab of Service Software and initialize the Liquid, Vacuum Cooling System.
- Click the On buttons for Cooling, Circulating Pump, and Additional Fan.
Ensure that the Reagent and Sample Bay temperatures come into range after a few minutes.
- Ensure that there are no leaks in the cooling system, especially around the right side of the Sample Bay and the rear of the Reagent Bay.
- Run a Fusion Transfer Test.
Service Procedures > Fusion Transfer Test
- Run a Side Car Pipettor OQ Test.
Service Procedures > Side Car Pipettor OQ Test
- Run a Panther Fusion System OQ Test.
Service Procedures > Panther Fusion System OQ Test
- Run a Fusion Pipettor Pressure Integrity Test.
Service Procedures > Fusion Pipettor Pressure Integrity Test
- Run Thermocycler Scanning and Automatic Cleaning
Click the