Flashcheck - Lumo Injector Verification Procedure

Purpose

To verify the Panther Detection System operation is in compliance with performance validity and acceptance criteria.

Scope

The Flashcheck Reagent is intended to be used for performing a Quality Control verification test of the Auto Detect fluid pumps and injectors of the Panther Luminometers as part of new system manufacturing and/or existing system service/maintenance procedures.
The use of the Flashcheck Reagent effectively assesses the Panther System detection injection system.

A Flashcheck is required for all Panther installs if the customer will run TMAClosedTranscription-mediated amplification—Hologic's patented nucleic acid amplification technology. assays.

If installing a Panther that is ONLY going to run Real Time Assays or installing a Panther Fusion that will ONLY run PCR assays, then a Flashcheck is NOT required on installation.

Time Required

  • 45 minutes

Flashcheck FSE Kit

ClosedRequired Tools and Materials

Description Quantity
Flashcheck Reagent FSE Kit (Refer to the Flashcheck FSE Kit package insert) 1
Empty Sample Tubes 4
DI water  
Note—See the Storage Instructions section in the Flashcheck FSE Kit package insert.

APTIMA Assay Fluids

Reagents Material Number
Auto Detect 1 LR0167G-01 or equivalent
Auto Detect 2 LR0148G-01 or equivalent
Oil LR0354 or equivalent
Wash Solution LR0356 or equivalent
Buffer for Deactivation Solution LR0355 or equivalent

ClosedPreparation

  1. ClosedPrepare 4 new empty sample tubes:
    1. Place tape or a label on one side of each tube.
    2. Place the four taped sample tubes in the first four positions of a sample rack. Ensure the tape on each sample tube is blocking the barcode on the inside of each sample rack slot.
Note—Alternatively, if a Syscheck was just completed, the four Syscheck Reagent tubes could be used here as long as the barcodes are facing inward.

  1. Prepare the reagent bottles according to the Flashcheck Reagent FSE Kit Reagent Preparation section of the Flashcheck Reagent FSE Kit package insert.
Note—This bottle needs to be reconstituted. Ensure there are no bubbles.
  1. ClosedPlace the bottles in a Reagent Rack according to the following sequence:
    (The lids are removed prior to loading in the Reagent Bay.)
  1. Amplification Reagent position
  2. Enzyme Reagent position
  3. Probe Reagent position
  4. Select Reagent position

ClosedProcedure

  1. Power on the Panther System and PC.
  2. Start Panther Main.
  3. Disable LIS Communication.
  1. Display the customer configuration settings by selecting the Admin tab, then System Configuration, then the drop-down menu for Specimen Loading.

  2. ClosedRecord the customer's settings (checked or unchecked) for the following options:
  • Enable Non-Barcode Sample Loading
  • Enable Auto Sample ID Assignment for Unreadable Barcodes
  • Enable Detection and Flagging of Repeat Barcodes
  1. ClosedChange the customer’s configuration settings to run Flashcheck:
  • CHECK Enable Non-Barcode Sampling Loading
  • UNCHECK Enable Auto Sample ID Assignment for Unreadable Barcodes
  • UNCHECK Enable Detection and Flagging of Repeat Barcodes

  1. Click Save.
  2. Navigate to the Tasks screen and perform the following Tasks according to the Panther System Operator's Manual.
  1. Load Tips
  2. Load MTUs
  3. Load Universal Fluids
  4. Empty Waste
  1. Click the PrimeClosedOperation of pumping fluid through tubing to ensure proper and consistent fluid delivery (remove air from the tubing, etc.). Instrument button. Wait for the Prime to complete successfully.
  2. ClosedClick the Load Assay Reagents button.

  1. Open the Reagent Bay door, and Closedwhen the lane indicator turns purple, load the Reagent Rack into the first lane of the Reagent Bay. Ensure the barcodes are visible and the all the lids are removed.

  1. Close the Reagent Bay door, and Closedverify the system recognizes the Flashcheck kit has been loaded.

  1. Open the TCR door and place the empty, labeled TCR bottle in the TCR Carousel Slot 1 (use a TCR Carousel adapter if necessary) with the barcode visible. The Reagent Bay lane and the TCR slot have to match.
  1. Shut the TCR door, then wait for the Carousel to scan and the TCR indicator to turn green.
  1. The reconstitution screen is displayed when the scan is successful.
    1. Click on the Calendar icon to enter the current date and click OK.
    2. Now hit Accept. This will take put you back on the Assay Reagent Screen.
  1. From the Tasks screen, click Load Samples.

The Sample Rack Bay screen appears.

  1. ClosedClick Set on the top right side of the screen.
  1. ClosedSelect the Replicates box.
  1. ClosedEnter 5 in the Replicates field and click OK. This creates 20 test orders with 4 sample tubes for the Flashcheck sequence.
  1. ClosedClick Add.
  1. ClosedSelect Flashcheck from the assay list and click OK.
  1. ClosedClick Accept.
  1. Ensure a sample rack lid is installed on the rack with the 4 sample test tubes (with blank labels or tape applied).
  Note—If you have just performed a Syscheck, you can reuse the rack with the 4 SysCheck reagent tubes, just turn the barcodes so they are not visible.
  1. Open the Sample Bay door and wait for Closedthe lane indicator to turn purple.

  1. Load the Sample Rack into the first lane and close the door.
  2. Verify the Sample Rack indicators turn green.
  1. ClosedVerify that 20 test orders were created for this sequence. by navigating to the TEST ORDERS tab. In the bottom left corner you will find total test orders.
  Note—You may have to select “By Date” to see the Test Orders instead of having “Not processed today” selected.
  1. The Flashcheck Reagent sequence begins automatically.
  Note—If the sequence does not begin after a few minutes, verify all the items in the Task screen have been highlighted green.
  1. The Flashcheck sequence takes approximately 20 minutes to complete.
  1. Enable LIS Communication.
  2. Set Sample Handling Configuration to customer's configuration recorded in Step 4.
  3. Proceed to General Data Analysis and Result Interpretation

ClosedGeneral Data Analysis and Result Interpretation

  1. Data Analysis and Result Interpretation via Flashcheck Spreadsheet
  2. ClosedData Analysis and Result Interpretation via Panther Remote Dashboard
  1. Launch the Panther Dashboard software.
  1. ClosedNavigate to the Tasks screen.
  1. ClosedClick the Perform Maintenance button.
  1. ClosedClick the Service button.
  1. ClosedClick the Dashboard button.
  1. ClosedClick the Flashcheck tab.
  1. ClosedEnter the Operator Name, Reagent Lot Number, and Expiration Date.
  1. If the system runs HPV or Parvo, Closedselect YES from the HPV USED dropdown menu. Select NO if neither HPV nor Parvo is run on this system.

  1. ClosedSelect the data set to be analyzed.
  1. ClosedClick Perform Calculation.
    A Flashcheck report is generated containing the pass/fail criteria for the run.
  1. Save Flashcheck report and attach to applicable service record.

ClosedAcceptance and Validity Criteria

Note—All the following Acceptance Criteria and Run Validity are presented by the Dashboard software Flashcheck application.

Run Validity

If one outlier is identified by the Dashboard software Flascheck process, the system will automatically remove that outlier and recalculate the Flashcheck Reagent results.
If the resulting calculation still fails, then the final determination will be failed.

If using the manual spreadsheet, the formulas in the cells will indicate pass/fail.

Acceptance Criteria

Non-HPV/Non-Parvo Systems

  • The luminometer should produce a valid run with data from the Flashcheck Worksheet indicating all of the following results:
  • Average fraction of Total RLUClosedRelative light unit at interval 3 ≥ 0.00500
  • Average fraction of Total RLU at interval 25 ≤ 0.00140
  • %CV at Interval 25 ≤ 100
  • Highest %CV of intervals 26 through 50 ≤ 120
  • For the pumps and injectors to be acceptable:
    • Average fraction of the Total RLU at interval 3 ≥ 0.00500
    • Average fraction (based on 4 replicates) of Total RLU at interval 25 ≤ 0.00140
  • The Panther Luminometer Flashcheck Reagent Verification procedure is considered PASS if each individual Result Interpretation on the Spreadsheet passes. A “FAIL” result is not acceptable.

HPV/Parvo Systems

  • The luminometer should produce a valid run with data from the Flashcheck Worksheet indicating all of the following results:
  • Average fraction of Total RLU at interval 3 ≥ 0.100
  • Average fraction of Total RLU at interval 25 ≤ 0.00070
  • %CV at Interval 25 ≤ 100
  • Highest %CV of intervals 26 through 50 ≤ 120
  • For the pumps and injectors to be acceptable:
    • Average fraction of the Total RLU at interval 3 ≥ 0.10
    • Average fraction (based on 20 replicates) of Total RLU at interval 25 ≤ 0.00070
  • The Panther Luminometer Flashcheck Reagent Verification procedure is considered PASS if each individual Result Interpretation on the Spreadsheet passes. A “FAIL” result is not acceptable.

Retest Criteria

  • If the initial test results do not meet acceptance criteria, repeat the Panther Luminometer FlashcheckReagent Verification.
    Repeat runs must meet acceptance criteria.
    Document any and all failed and passed Flashcheck Procedures in the applicable service
Note—Failure to meet acceptance criteria after repeat performance may indicate faulty performance of equipment. If necessary, replace the suspected faulty equipment, repeat performance of the Flascheck verification test, and evaluate results.