Draining the Coalescing Trap Procedure
Parts and Materials Required
- Proper PPE
- Flat tray to catch fluid
- Allen wrench, 8 mm
Time Required
- 10 minutes
Removal Procedure
- Put on proper PPE.
- Open the Waste Drawer.
Locate the coalescing drain plug on the right side of the Waste Drawer.- Place a flat tray underneath the coalescing drain plug to catch any fluid that drains out.
Using an 8 mm Allen wrench, slowly open the coalescing drain plug.
After all fluid has drained, close the coalescing drain plug.- Properly dispose of the liquid waste fluid from the drain.
- Clean the areas with bleach/water/alcohol that fluid from the trap came into contact with.
- Properly dispose of gloves.