Unloading Undeactivated MTUs from Mag Parking, Sample Dispense Slot, Chiller Ramp, or Sample Mix Station

Purpose

To explain the method for removing and deactivating MTUs from Mag Parking, Sample Dispense Slot, Chiller Ramp, or Sample Mix Station in the event of a failure.

Parts and Materials Required

Procedure

  1. Put on proper PPE.
  2. Power down the Panther System.
  3. Dispose of MTUClosedMulti-tube unit—Container used to process tests in the instrument. An MTU contains five separate reaction tubes. The MTU is moved through the instrument by the linear distributor and includes five tiplets for pipettiing to be used in the mag wash station. waste bag and replace with a new waste bag.
  4. Change Gloves.
  5. Open the right and left side panel doors.
  6. Verify that both pipettor arms are up. Manually push them up if they are not.
  7. ClosedRemove the Luminometer injectors by unscrewing the two screws that secure the injectors to the luminometer module.

  8. Unlock the Mid Bay drawers (left and right side) and pull the Mid Bay all the way out.
  9. ClosedSlowly pull the MTU out of the slot containing the MTU.

  10. Add bleach to each tube as the MTU is being pulled out.
  11. Gently place the MTU in the waste bin.
  12. Wait 15 minutes then dispose the waste bag in a bio-hazard waste container.