Unloading Undeactivated MTUs Luminometer
Purpose
To explain the method for removing and deactivating MTUs from the Luminometer in the event of a failure.
Parts and Materials Required
- Proper PPE
- Panther Tool Kit
- Squirt Bottle with recommended sodium hypochlorite concentration
- Procleix Assays: 0.5 to 0.7%
- Aptima
Automated Probe Transcription Mediated Amplification. Trade name for amplified NAT infectious disease assays and reagents designed for use with Hologic's instrument platforms. and other diagnostic assays: 2.5 to 3.5%
- Absorbent Bench Pads
- Hex Wrench Set
Procedure
- Put on proper PPE.
- Power down the Panther System.
- Dispose of MTU
Multi-tube unit—Container used to process tests in the instrument. An MTU contains five separate reaction tubes. The MTU is moved through the instrument by the linear distributor and includes five tiplets for pipettiing to be used in the mag wash station. waste bag and replace with a new waste bag. - Change Gloves
- Open the right and left side panel doors.
- Verify that both pipettor arms are up. Manually push them up if they are not.
Remove the Luminometer injectors by unscrewing the two screws that secure the injectors to the luminometer module.
- Place the injectors securely in the injector slot on the TCR
Target capture reagent—An assay-specific reagent added as part of specimen pipetting. module.
Note—Older style TCR modules will not have this place holder. In that case, place injectors on an absorbent pad or wrap it with an absorbent towel. - Unlock the Mid Bay drawers (left and right side) and pull the Mid Bay all the way out.
Remove the 10 screws located on top of the Luminometer cover.
- Carefully lift the cover off the Luminometer and rest the cover on top of a blue pad to the side of the Luminometer to allow access to the MTU.
To open the door, use a long hex screw driver and apply pressure to the locking mechanism and push it downward, this will open the Luminometer door.
Slide the transport drive mechanism to the right.
Slide the MTU out of the Luminometer one tube at a time, dispensing bleach from a bleach bottle into each tube.
Slide in the Mid Bay drawer and pull out the waster drawer and gently place the deactivated MTU into the waste drawer, being careful not to sill the contents of the MTU.
- Change gloves.
- Wait 15 minutes to ensure complete deactivation of the MTU and then remove waste bag and dispose in a bio-hazard waste container.