Unloading Undeactivated MTUs from Amp and HPA Load Station

Purpose

To explain the method for removing and deactivating MTUs from the Amp and HPAClosedHybridization protection assay—The Hologic HPA technique uses a specific DNA probe, labeled with an acridinium ester detector molecule that emits a chemiluminescent signal. Load Station in the event of a failure.

Parts and Materials Required

Procedure

  1. Put on proper PPE.
  2. Power down the Panther System.
  3. Dispose of MTUClosedMulti-tube unit—Container used to process tests in the instrument. An MTU contains five separate reaction tubes. The MTU is moved through the instrument by the linear distributor and includes five tiplets for pipettiing to be used in the mag wash station. waste bag and replace with a new waste bag.
  4. Change Gloves.
  5. Open the right and left side panel doors.
  6. Verify that both pipettor arms are up. Manually push them up if they are not.
  7. ClosedRemove the Luminometer injectors by unscrewing the two screws that secure the injectors to the luminometer module.

  8. Unlock the Mid Bay drawers (left and right side) and pull the Mid Bay all the way out.
  9. ClosedSlide the load station door to the left to access MTU.

  10. ClosedPull the MTU slowly out of the load station.

  11. ClosedBleach each tube of the MTU as it is being pulled out of the load station.

  12. Gently place MTU into the waste bin.
  13. Wait 15 minutes and then dispose the waste bag in a bio-hazard waste container.