Unloading Undeactivated MTUs from Linear Distributor
Purpose
To explain the method for removing and deactivating MTUs from the Linear Distributor in the event of a failure.
Parts and Materials Required
- Proper PPE
- Panther Tool Kit
- Squirt Bottle with recommended sodium hypochlorite concentration
- Procleix Assays: 0.5 to 0.7%
- Aptima
Automated Probe Transcription Mediated Amplification. Trade name for amplified NAT infectious disease assays and reagents designed for use with Hologic's instrument platforms. and other diagnostic assays: 2.5 to 3.5%
- Absorbent Bench Pads
- Hex Wrench Set
Procedure
- Put on proper PPE.
- Power down the Panther System.
- Dispose of MTU
Multi-tube unit—Container used to process tests in the instrument. An MTU contains five separate reaction tubes. The MTU is moved through the instrument by the linear distributor and includes five tiplets for pipettiing to be used in the mag wash station. waste bag and replace with a new waste bag. - Change Gloves
- Open the right and left side panel doors.
- Verify that both pipettor arms are up. Manually push them up if they are not.
Remove the Luminometer injectors by unscrewing the two screws that secure the injectors to the luminometer module.
- Unlock the Mid Bay drawers (left and right side) and pull the Mid Bay all the way out.
- Slide the distributor all the way to the left side of the system.
Turn the distributor counter-clockwise until the MTU is facing outward and visible.
Carefully slide the MTU out of the distributor head one tube at a time.
Caution— Dispense bleach into each tube as they are removed from the distributor.
With the MTU filled with bleach, carefully open the waste drawer, and place the MTU inside the waste bag.
- Change gloves.
- Wait 15 minutes then dispose the waste bag in a bio-hazard waste container.