Servicing and Replacing Fluid in the Liquid Cooling Module
Parts and Materials Required
- 2L empty fluid container(s)
- Absorbent pad
- Cooling Module Fill Tube Tubing Assembly
- Proper PPE
Time Required
Procedure
| |
|
Put on proper PPE. |
- Drain the Cooling Module.
- Service the Cooling Module.
- Follow the instructions for Filling the Cooling Module.
- Follow the instructions for Draining the Cooling module.
- Repeat steps 2a and 2b until no foreign particulates or discoloration is observed in the DI water extracted in step 2b.
- Fill the Cooling Module.
Verification
- Verify the presence of fluid in the Cooling Module fluid line and reservoir.
- Visually verify the absence of foreign particles and discoloration in the Cooling Module fluid line.
- Visually verify that there are no leaks from the Cooling Module or any of the fluid connections.
- Verify that the Reagent and Sample Bays reach their specified temperatures.
Click the